• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC 150MM, STERILE, PERC REF PIN; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC 150MM, STERILE, PERC REF PIN; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9733236
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Unspecified Tissue Injury (4559)
Date of Event 02/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
No devices were returned to medtronic for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation device being used intra/peri-operatively of a sacroiliac and thoracolumbar procedure.It was reported that, while malleting the perc pin and it broken while still inside the patient.There was a less than 1-hour delay to the procedure and no impact on patient outcome.The likely cause of the issue was the perc pin being hammered into a previously applied hardware.The surgeon decided to leave the broken piece inside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
150MM, STERILE, PERC REF PIN
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key11401565
Report Number1723170-2021-00546
Device Sequence Number4760364
Product Code OLO
UDI-Device Identifier00613994247865
UDI-Public00613994247865
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K131425
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2023
Device Model Number9733236
Device Catalogue Number9733236
Device Lot Number2020110320
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2021
Initial Report FDA Received Date03/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient Weight57
-
-