• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS-70MM; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS-70MM; GAUGE, DEPTH Back to Search Results
Model Number SD319.002
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a j&j sales representative.The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the four depth gauge sleeves were found after sets had been through decontamination and the inner part of the depth gauge had been disposed of by central sterile staff due to the inner part of the depth gauge being bent or broken.There was no patient involvement.This report is for one (1) depth gauge for 2.0mm and 2.4mm screws-70mm.This is report 1 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a product investigation was conducted.Visual inspection: the depth gauge for 2.0mm and 2.4mm screws-70mm (p/n: sd319.002, lot number: 6504650) was received at us cq.Visual inspection of the complaint device showed the slider component was not returned due to being disposed of by central sterile staff.The returned body component had no observed damage.Dimensional inspection: a dimensional inspection was not performed since it was not applicable to the complaint condition.Document/specification review: relevant drawings current and (manufactured)were reviewed.No design issues or discrepancies were identified.Investigation conclusion: this complaint is not confirmed as the returned part component was not damaged and the other components were disposed of and not missing.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: part # sd319.002, synthes lot # 6504650, supplier lot # na, release to warehouse date: 13 may 2011, manufactured by synthes brandywine.No ncr's were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS-70MM
Common Device Name
GAUGE, DEPTH
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11401715
Report Number8030965-2021-01529
Device Sequence Number19966537
Product Code HTJ
UDI-Device Identifier10886982004925
UDI-Public(01)10886982004925
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSD319.002
Device Catalogue NumberSD319.002
Device Lot Number6504650
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2021
Initial Date Received by Manufacturer 02/07/2021
Supplement Date Received by Manufacturer03/15/2021
Initial Report FDA Received Date03/02/2021
Supplement Report FDA Received Date03/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-