• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 7-8 11MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN PS HIGH FLEX XLPE SZ 7-8 11MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71453232
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a tka procedure insert would not lock into tibia tray.A backup item was opened and seated correctly.Instruments were inside the patient.There was no delay or further complications.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned insert confirms minor damage to the lock detail, more than likely from attempting to snap into the tibia tray.A dimensional inspection was attempted but the device was too damaged from the attempted insertion to obtain accurate measurements.A review of complaint history did not reveal additional complaints for the listed item.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 7-8 11MM
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11403057
Report Number1020279-2021-01771
Device Sequence Number4568644
Product Code JWH
UDI-Device Identifier00885556035153
UDI-Public00885556035153
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/15/2021
Device Model Number71453232
Device Catalogue Number71453232
Device Lot Number11MM09572
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2021
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer06/17/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date06/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-