• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DURALOC CONST LINER 48X28; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DURALOC CONST LINER 48X28; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1241-08-527
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient dislocated.Removed depuy duraloc cup and duraloc constrained liner and 28mm ts head.This is patients 2nd or 3rd revision.Reimplanted with competitor cup and constrained liner.Doi: (b)(6) 2020, dor: (b)(6) 2021, affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information was received stating that the patient had a previous surgery where a constrained duraloc liner was implanted.That¿s the only time they were aware of involving depuy products for dislocation.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURALOC CONST LINER 48X28
Common Device Name
DURALOC IMPLANT : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11404150
Report Number1818910-2021-04199
Device Sequence Number4503240
Product Code KWZ
UDI-Device Identifier10603295018612
UDI-Public10603295018612
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P960054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1241-08-527
Device Catalogue Number124108527
Device Lot NumberJ0660X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/17/2021
Supplement Date Received by Manufacturer02/17/2021
03/31/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date03/17/2021
04/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DLT TS CER HD 12/14 28MM +8.5; DURALOC DYNAMIC LOCK RING 48MM; PINN CAN BONE SCREW 6.5MMX45MM; DLT TS CER HD 12/14 28MM +8.5; DURALOC DYNAMIC LOCK RING 48MM; PINN CAN BONE SCREW 6.5MMX45MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight45
-
-