EXACTECH, INC. EQUINOXE; HUMERAL STEM PRIMARY, PRESS FIT 15MM Back to Search Results
Model Number 300-01-15
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
As reported, during impaction phase of the procedure on this (b)(6) y/o male patient, the implant, broach trial, and reamer for a size 15 did not match sizes and surgeon ended up having a subsided stem."specifically trial and real implant." the surgeon had to ream the canal again and broach up a size in order to get implant to fit.Created a stressful situation for surgeon.This created a short delay with no harm to the patient.Patient was last known to be in stable condition following the event.Device to return.
 
Additional Manufacturer Narrative
Section h10: h3: the size discrepancy reported was likely the result of the stem trail's distal diameter being manufactured at mmc and proximal m/l width being undersized while the final implant's distal diameter being manufactured closer to lmc, which led to the stem trial sitting proud and the final implant subsiding.Additional contributing factor may include user error or over impacting when broaching the humerus or poor bone quality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
HUMERAL STEM PRIMARY, PRESS FIT 15MM
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key11404350
Report Number1038671-2021-00084
Device Sequence Number14338898
Product Code KWT
UDI-Device Identifier10885862079336
UDI-Public10885862079336
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K042021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number300-01-15
Device Catalogue Number300-01-15
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/11/2021
Supplement Date Received by Manufacturer08/10/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date08/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age76 YR
-
-