| Catalog Number |
72204114 |
| Medical Device Problem Codes |
Failure to Shut Off (2939); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a foot surgery, the forward button of the dyonics handpiece was sticky but when it was released the handpiece did not shut off.The procedure was successfully completed without delay using a back-up device.No patient injury or complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: case (b)(4).The reported device was received for evaluation.A visual inspection found cosmetic scratches, and the forward trigger stuck.A functional evaluation was performed by the supplier.The device failed testing due to the trigger is stuck in the housing.The device was inspected and found nicks in the trigger housing causing the trigger to get stuck, also the controller failed for high current and the lohet in the base is bad.It was determined the device did not contribute to the reported event.The complaint was confirmed and the root cause was associated with unintentional use of the device.Factors that could have contributed to the reported event include an impact to the device and improper maintenance.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.A review of risk management files found that the reported failure was documented appropriately.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|