• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US GLOBAL SHD END PEG GLEN 52; GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US GLOBAL SHD END PEG GLEN 52; GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID Back to Search Results
Model Number 1137-87-025
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision of anatomic shoulder to reverse shoulder - aseptic loosening of anatomic shoulder components, implants removed and reverse unable to be completed, therefore, reverse stem with cta head was implanted.The head was removed with the removal tool, then both the stem and the glenoid were able to be removed with little to no effort - were more just sitting in the humeral canal and glenoid bone rather than attached to these interfaces.
 
Event or Problem Description
Additional information received indicated that the affected side was the right shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : no anomalies or deviations were found during the review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLOBAL SHD END PEG GLEN 52
Common Device Name
GLENOID (KEELED AND 5-PEGGED) IMPLANT : SHOULDER GLENOID
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11405242
Report Number1818910-2021-04262
Device Sequence Number11895139
Product Code KWT
UDI-Device Identifier10603295007289
UDI-Public10603295007289
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1137-87-025
Device Catalogue Number113787025
Device Lot NumberXB4KN1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/17/2021
Supplement Date Received by Manufacturer04/29/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date04/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLOBAL ADVANTAGE ECC HD 52X21; GLOBAL ADVANTAGE STEM 10 MM; GLOBAL SHD END PEG GLEN 52; GLOBAL ADVANTAGE ECC HD 52X21; GLOBAL ADVANTAGE STEM 10 MM; GLOBAL SHD END PEG GLEN 52
Outcome Attributed to Adverse Event Required Intervention;
-
-