DEPUY ORTHOPAEDICS INC US PINN LNR CON +4 10D 32IDX54OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS Back to Search Results
Model Number 1218-32-754
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Illegible Information (4050)
Health Effect - Clinical Codes Joint Dislocation (2374); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the original implant date is unknown.The doctor said the patient dislocated their prosthetic hip.The constraining ring was loose from the liner and showed signs of impingement (dented).Poly and head were removed and replaced.Lot numbers were unreadable upon removal.Doi: unknown, dor: (b)(6) 2021, affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a device history record (mre) review, was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN LNR CON +4 10D 32IDX54OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11405491
Report Number1818910-2021-04285
Device Sequence Number14585866
Product Code KWZ
UDI-Device Identifier10603295012009
UDI-Public10603295012009
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-32-754
Device Catalogue Number121832754
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/10/2021
Supplement Date Received by Manufacturer03/23/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date03/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 32 +5 BR; PINN LNR CON +4 10D 32IDX54OD; ARTICUL/EZE BALL 32 +5 BR; PINN LNR CON +4 10D 32IDX54OD
Outcome Attributed to Adverse Event Required Intervention;
-
-