| Model Number |
1218-32-754 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Illegible Information (4050)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/10/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the original implant date is unknown.The doctor said the patient dislocated their prosthetic hip.The constraining ring was loose from the liner and showed signs of impingement (dented).Poly and head were removed and replaced.Lot numbers were unreadable upon removal.Doi: unknown, dor: (b)(6) 2021, affected side: left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a device history record (mre) review, was not possible because the required lot code was not provided.
|
| |
|
Search Alerts/Recalls
|