Model Number 1219-36-154
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 08/07/2013
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision total hip replacement.Reason for revision: instability and dislocation.Primary surgery carried out on (b)(6) 2008.A constrained pinnacle liner was implanted.Right side.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : no anomalies or deviations were found during the review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR +4 10D 36IDX54OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11407064
Report Number1818910-2021-04308
Device Sequence Number3545244
Product Code LPH
UDI-Device Identifier10603295014584
UDI-Public10603295014584
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K033273
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2013
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1219-36-154
Device Catalogue Number121936154
Device Lot NumberCC7DE1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/18/2021
Supplement Date Received by Manufacturer05/05/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date05/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELTA CER HEAD 12/14 36MM +1.5; PINN MAR +4 10D 36IDX54OD; DELTA CER HEAD 12/14 36MM +1.5; PINN MAR +4 10D 36IDX54OD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-