Zoll medical corporation evaluated the device and the electrodes and the device performed to specification.The device was recertified and returned to the customer.The device was put through extensive testing including bench handling, ecg monitoring stress testing, and defib cycle testing without duplicating any malfunction to the device.The pads that were used during the event were returned stuck together and were compromised but did pass continuity testing.Review of the log did see evidence of the reported problem but does not indicate a device malfunction.The log showed the device is charged but is in a defib lead fault state and energy cannot be delivered.A defib lead fault indicates that a valid patient impedance was not detected through the pads.There are multiple factors outside of a device malfunction that could contribute to this condition.Some include, but are not limited to motion artifacts, poor contact between the pads and the patient's skin, poor contact between the multifunction cable and the adapter, or the patient's skin condition.This report has been attributed to poor coupling of the electrode pads to the patient's skin.Analysis of reports of this type has not identified an increase in trend.
|