(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.The customer reported medication could not be injected into the catheter.The customer returned one flat filter, one snaplock assembly, and epidural catheter.The returned components were received connected together (reference attached files inp1900082908).The returned components were visually examined with and without magnification.Visual examination of the returned filter and snaplock assembly revealed both components appear typical with no observed defects or anomalies.Visual examination of the returned catheter revealed the catheter appears used.Biological material can be seen between the inner coils.There is a heavy presence of biological material at the distal end of the returned catheter.A functional flow test was performed on the returned sample per (b)(6) section 7.8; rev.8.The returned epidural catheter was inserted from the proximal end into the returned snaplock assembly until it bottomed out and the snaplock assembly was closed.The components were confirmed to be secured by tugging gently.The snaplock assembly was connected to the lab leak tester (b)(4) and the pressure was increased to 10 psi to establish flow.A very slow flow with reddish material could be established out of the distal tip of the catheter.The test was repeated using the returned snaplock assembly and lab inventory epidural catheter.Flow could be established to the distal tip of the catheter, indicating that there are no flow issues with the returned snaplock assembly.An attempt was made to thread a lab inventory wire through the epidural catheter from the proximal end.The wire threaded completely through the catheter from the proximal end of the returned catheter, but with some resistance.An attempt was made to thread a lab inventory wire through the epidural catheter from the distal end.The wire threaded completely through from the distal end of the returned catheter also with some resistance.The wire was able to push the occlusion on through the catheter.A flow test was once again performed with the returned catheter and returned snaplock assembly.Flow was established coming from the distal tip.However, the initial flow was reddish material that was pushed through and the flow became clear, indicating biological material present in the catheter was causing the occlusion.The flow rate was measured at 7.0ml/min (stopwatch: ref-003143), which is within the specification of a minimum 1ml/min.A corrective action is not required at this time as the condition of the sample received indicates unintentional user error caused or contributed to this event.The reported complaint of medication not injecting into the catheter was confirmed during functional inspection of the returned sample.The returned epidural catheter was found to be partially occluded with biological material.Microscopic examination revealed a heavy presence of biological material at the distal end.A device history record review was performed on the epidural catheter with no evidence to suggest a manufacturing related cause.After the lab inventory wire pushed the biological material through, the returned components passed a functional flow test and met flow rate specifications.Therefore, based upon the condition of the sample received, unintentional user error caused or contributed to this event.
|