| Model Number |
M00550310 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pulmonary dilatation balloon was used during a trachea dilation procedure performed on (b)(6) 2021.During the procedure, the balloon was inflated twice to 7atm.However, the balloon ruptured and deflated.It was noted that the pressure at the time of the burst was greater than 7atm.The procedure was completed with another cre pulmonary dilatation balloon device.There were no patient complications reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|