• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE FG., NANOTACK TT SUTURE ANCHOR, 1.4MM WITH 1.2MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE FG., NANOTACK TT SUTURE ANCHOR, 1.4MM WITH 1.2MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number CAT02969
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.All pieces were retrieved.
 
Event or Problem Description
It was reported that the anchor broke during the procedure.All pieces were retrieved.
 
Additional Manufacturer Narrative
Alleged failure: anchor broken.Probable root cause: design: inserter/implant not compatible with guide.Inserter material/design too weak.Inserter shaft cannot bend around curved guide.Guide mouth not designed to grip bone.Guide not able to be gripped tightly.Clearance between anchor coupling and inserter id too large.Inserter tube wall thickness too thin.Process inserter, implant, and/or guide manufactured out of spec.Anchor not assembled properly to inserter.Application: excessive force impacting inserter.Movement of guide out of orientation.Guide buried into bone.The product was not returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.H3 other text: 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FG., NANOTACK TT SUTURE ANCHOR, 1.4MM WITH 1.2MM XBRAID TT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11408096
Report Number0002936485-2021-00123
Device Sequence Number19380854
Product Code MBI
UDI-Device Identifier07613327294606
UDI-Public07613327294606
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K131769
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCAT02969
Device Catalogue NumberCAT02969
Device Lot Number20195AE2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/06/2021
Supplement Date Received by Manufacturer02/06/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date05/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-