| Model Number |
CAT02969 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure.All pieces were retrieved.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor broke during the procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: anchor broken.Probable root cause: design: inserter/implant not compatible with guide.Inserter material/design too weak.Inserter shaft cannot bend around curved guide.Guide mouth not designed to grip bone.Guide not able to be gripped tightly.Clearance between anchor coupling and inserter id too large.Inserter tube wall thickness too thin.Process inserter, implant, and/or guide manufactured out of spec.Anchor not assembled properly to inserter.Application: excessive force impacting inserter.Movement of guide out of orientation.Guide buried into bone.The product was not returned for investigation and the reported failure mode was not confirmed.The failure mode will be monitored for future reoccurrence.H3 other text: 81.
|
| |
|