• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. SENSAR IOL; INTRAOCULAR LENS Back to Search Results
Model Number AR40M
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.If explanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a sensar intraocular lens (iol) trailing haptic was kinked upon insertion into the patient's operative eye.A backup lens of the same model and diopter was used to complete the procedure.There was no additional surgical intervention performed and the patient is reportedly doing well post-operation.No further information provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes section d9: return to manufacturer: 3/16/2021.Section h3: device evaluated by manufacturer: yes device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during removal and replacement.Haptic damage (bent) was also observed.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue was confirmed, but can be attributed to handling and cannot be confirmed to be related to manufacturing and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
MDR Report Key11408601
Report Number2648035-2021-07330
Device Sequence Number3631186
Product Code HQL
UDI-Device Identifier05050474501805
UDI-Public(01)05050474501805(17)250916
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/16/2025
Device Model NumberAR40M
Device Catalogue NumberAR40MN0020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/03/2021
Supplement Date Received by Manufacturer04/11/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date05/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EMERALDC30 CARTRIDGE; EMERALDC30 CARTRIDGE
Patient Age35 YR
-
-