• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII OXINIUM FEM SIZE 7 LEFT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII OXINIUM FEM SIZE 7 LEFT; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71420146
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Loss of Range of Motion (2032)
Date of Event 02/08/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, patient had a total knee replacement 3 year ago.Patient suffered from a stiff knee and limited range of motion 20 degree - 90 degree and overstuffed patella femoral joint, therefore patient had a revision surgery, where surgeon removed the patella, femoral, tibial insert and tibial baseplate from primary surgery.Surgeon noticed the patella was not resected enough, which cause over stuffing.Surgeon achieved a range of motion from 0 degrees - 120 degrees range of motion.Patient outcome is unknown.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A medical investigation was conducted and confirms this case reports a revision surgery was performed due to limited rom and an overstuffed patella.Per email correspondence, the requested surgical reports and x-rays are not available.Therefore, the patient impact beyond the revision cannot be determined.Due to the limited information provided, a thorough assessment cannot be rendered.Should additional clinically relevant documentation become available, the clinical/medical task may be re-evaluated.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of risk management files and instructions for use found that the reported failure was documented appropriately.Some potential probable causes could be alignment, fit/size of device used or wear.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII OXINIUM FEM SIZE 7 LEFT
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11408614
Report Number1020279-2021-01782
Device Sequence Number19718913
Product Code JWH
UDI-Device Identifier03596010439413
UDI-Public03596010439413
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K021673
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71420146
Device Catalogue Number71420140
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/08/2021
Supplement Date Received by Manufacturer02/08/2021
03/29/2021
Initial Report FDA Received Date03/03/2021
Supplement Report FDA Received Date03/04/2021
04/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-