Model Number 71674558
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during procedure the lag screw length gauge ((b)(4)) and the sureshot hexdriver ((b)(4)) were noticed broken.The devices broke outside the patient.No delay or injury reported.The procedure was finished using a smith and nephew backup device.
 
Additional Manufacturer Narrative
The devices, used in treatment, were returned and evaluated.A visual inspection of the returned devices did confirm they were broken and appear to be excessively worn.The manufactured date for this device is 2017.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.These devices are reusable instruments that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.D9 and h3: returned device b5: description update.
 
Event or Problem Description
It was reported that during procedure the lag screw length gauge (case-2021-00040944-1) and the sureshot hexdriver (b)(4) were noticed broken.The devices broke outside the patient.No delay or injury reported.The procedure was finished using a smith and nephew backup device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW LENGTH GAUGE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11408833
Report Number1020279-2021-01797
Device Sequence Number14338789
Product Code HSB
UDI-Device Identifier00885556363393
UDI-Public00885556363393
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71674558
Device Catalogue Number71674558
Device Lot Number17AM14822
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/08/2021
Supplement Date Received by Manufacturer03/26/2021
Initial Report FDA Received Date03/04/2021
Supplement Report FDA Received Date04/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-