| Model Number |
7213 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a balloon leak occurred.A 3.75mm x 15mm nc emerge balloon catheter was advanced for dilation.However, during procedure, it was noted that the balloon failed to hold pressure and it was leaking fluid.The device was removed and the procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon leak occurred.A 3.75mm x 15mm nc emerge balloon catheter was advanced for dilation.However, during procedure, it was noted that the balloon failed to hold pressure and it was leaking fluid.The device was removed and the procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was contrast and blood in the inflation lumen and balloon.The balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.There was a pinhole at the distal marker band.There was no marker band damage detected.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported balloon leak.
|
| |
|