• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG DAA DOUBLE OFFSET ADAPTER LEFT 80/45; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG DAA DOUBLE OFFSET ADAPTER LEFT 80/45; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 75102237
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the broach handle is loose and will not hold the trial firmly.The failure in the device was noticed during set up or inspection.The procedure was completed using the same device.No surgical delay or injury to the patient was reported.
 
Additional Manufacturer Narrative
Additional information in d4, g4, h4, h6 investigation results: it was reported that the broach handle is loose and will not hold the rasp firmly.No surgical delay or injury to the patient was reported.The device, intended for use in treatment, was returned for investigation.Upon visual inspection, the reported failure mode could not be confirmed.The device shows very limited signs of use but no damage is observed on the connector site to the broach.Furthermore, the instrument was tested with gauge id-nr 0723.No deviation was detected.No additional complaint with batch a60410 was reported so far.A review of the production documentation for batch a60410 did not reveal any deviation from the standard operating procedure.Based on the performed investigations, the reported failure mode could not be confirmed.However, previous investigations demonstrated that under specific circumstances the double offset adapter 80/45 may disconnect from the rasp during backslapping.An optimized design of the device has been released in order to reduce the occurrence of this issue.This version of the device will be monitored for similar issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAA DOUBLE OFFSET ADAPTER LEFT 80/45
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
MDR Report Key11409069
Report Number9613369-2021-00089
Device Sequence Number3853390
Product Code JDI
UDI-Device Identifier00885556535424
UDI-Public00885556535424
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75102237
Device Catalogue Number75102237
Device Lot NumberA60410
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/08/2021
Supplement Date Received by Manufacturer03/29/2021
Initial Report FDA Received Date03/04/2021
Supplement Report FDA Received Date04/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-