• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MBT REVISION CEM TIB TRAY SZ 6; MBT TIBIAL TRAY : KNEE TIBIAL TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MBT REVISION CEM TIB TRAY SZ 6; MBT TIBIAL TRAY : KNEE TIBIAL TRAY Back to Search Results
Model Number 1294-35-160
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Perforation of Vessels (2135); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/08/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Clinical notification received for revision of left total knee to address ongoing recalcitrant infection, and dysvascular left lower extremity with tissue loss date of implantation: (b)(6) 2016.Date of revision #1: (b)(6) 2017 (insert exchanged).Date of revision #2: (b)(6) 2017.(left knee).Treatment: left above the knee amputation (so as a consequence, effectively all implants removed).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Removed impact code device explantation (f1903).
 
Event or Problem Description
On (b)(6) 2016, the patient had a left cemented total knee arthroplasty to address primary end-stage left knee, severe with severe varus deformity and medial tibial bone loss.Contractures were noted during the procedure.On (b)(6) 2017, the patient had a revision left total knee tibial components to address infected left total knee arthroplasty secondary to contiguous infection and breakdown of left total knee arthroplasty wound.Only the insert was revised.On (b)(6) 2017, the patient had a left above the knee amputation to address chronic recalcitrant left total knee arthroplasty infection.Gross pus was found within the knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REVISION CEM TIB TRAY SZ 6
Common Device Name
MBT TIBIAL TRAY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11413034
Report Number1818910-2021-04367
Device Sequence Number3471080
Product Code NJL
UDI-Device Identifier10603295025993
UDI-Public10603295025993
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1294-35-160
Device Catalogue Number129435160
Device Lot Number412055
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/22/2021
Supplement Date Received by Manufacturer03/01/2021
04/19/2021
Initial Report FDA Received Date03/04/2021
Supplement Report FDA Received Date03/25/2021
04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPUY CMW 3 40G; DEPUY CMW 3 40G; MBT REVISION CEM TIB TRAY SZ 6; PFC*SIGMA/OV/DOME PAT 3PEG,41; SIG RP AOX STB INS SZ5 25; SIGMA PS CEM FEM SZ5 L; DEPUY CMW 3 40G; DEPUY CMW 3 40G; MBT REVISION CEM TIB TRAY SZ 6; PFC*SIGMA/OV/DOME PAT 3PEG,41; SIG RP AOX STB INS SZ5 25; SIGMA PS CEM FEM SZ5 L
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient Weight100
-
-