| Model Number |
1960-40-500 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Perforation of Vessels (2135); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/08/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Clinical notification received for revision of left total knee to address ongoing.Recalcitrant infection, and dysvascular left lower extremity with tissue loss.Date of implantation: (b)(6) 2016.Date of revision # 1: (b)(6) 2017 (insert exchanged).Date of revision # 2: (b)(6) 2017; (left knee).Treatment: left above the knee amputation (so as a consequence, effectively all implants removed).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, the patient had a left cemented total knee arthroplasty to address primary end-stage left knee, severe with severe varus deformity and medial tibial bone loss.Contractures were noted during the procedure.On (b)(6) 2017, the patient had a revision left total knee tibial components to address infected left total knee arthroplasty secondary to contiguous infection and breakdown of left total knee arthroplasty wound.Only the insert was revised.On (b)(6) 2017, the patient had a left above the knee amputation to address chronic recalcitrant left total knee arthroplasty infection.Gross pus was found within the knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Removed impact code device explantation (f1903).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|