• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY Back to Search Results
Model Number PXSLIMLAN150STR
Medical Device Problem Codes Material Deformation (2976); Physical Resistance/Sticking (4012)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the renal artery using a lantern delivery microcatheter (lantern) and ruby coils.During the procedure, the physician used a lantern to navigate the vessel and performed an angiogram.Later, while repositioning the lantern for a third time, the physician experienced resistance.Subsequently, the physician kinked the proximal end of the lantern.Therefore, the lantern was removed.The procedure was completed using ruby coils and a new lantern.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned lantern confirmed that the catheter was kinked at the proximal end.If the lantern is forcefully mishandled at an extreme angle during use, damage such as a kink may occur.Further evaluation revealed additional kinks and fractures throughout the length of the catheter shaft.This damage was incidental to the reported complaint and may have occurred during packaging of the device for return.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11413288
Report Number3005168196-2021-00397
Device Sequence Number14545369
Product Code DQY
UDI-Device Identifier00814548016672
UDI-Public00814548016672
Combination Product (Y/N)Y
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/05/2022
Device Model NumberPXSLIMLAN150STR
Device Catalogue NumberPXSLIMLAN150STR
Device Lot NumberF94074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/02/2021
Supplement Date Received by Manufacturer05/20/2021
Initial Report FDA Received Date03/04/2021
Supplement Report FDA Received Date05/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
Date Report Sent to FDA01/01/2005
-
-