| Model Number |
PFSR110006 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The navio long attachment used for treatment was returned for evaluation.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The evaluation followed the long attachment performance verification.The reported problem was confirmed.Internal bearings are disintegrated causing an obstruction in the long attachment.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints associated with the part with the failure modes ¿difficulty to insert", "break", "break, unintended movement", and "jamming seizing" identified similar events.The most likely cause of this event is degradation/collapse of the long attachment internal bearings making insertion difficult or not possible.Further investigation into the reported failure is being conducted to determine if additional actions are required.Our reference number: case-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a navio ukr procedure, the long attachment would not let burr go through it.It was swapped for its backup to continue with a delay of fewer than 30 minutes.No other complications were reported.It was found during investigation that the internal bearings are disintegrated causing an obstruction.
|
| |
|