U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST60; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX ST60; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 115308
Device Problems Leak/Splash (1354); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/10/2021
Event Type  malfunction  
Manufacturer Narrative
Eprismaflex st60 set ckt has been temporarily approved for use in the us under emergency use authorization (b)(4) to deliver crrt to treat patients in an acute care environment during the covid-19 pandemic.Initial reporter phone number: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported during treatment with a prismaflex st60 set, an external blood leak was observed due to a torn line.The amount of blood loss was reported as 130 ml.The treatment was immediately stopped.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
H10: the device was not received for evaluation, however a picture was provided.The visual inspection observed an external fluid leak from the access line multi connector.The reported condition was verified.The cause was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX ST60
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11415257
MDR Text Key244521170
Report Number8010182-2021-00089
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414118372
UDI-Public(01)07332414118372
Combination Product (y/n)Y
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Catalogue Number115308
Device Lot Number20G2207A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/10/2021
Initial Date FDA Received03/04/2021
Supplement Dates Manufacturer Received04/07/2021
Supplement Dates FDA Received04/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-