| Model Number |
ZA9003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If implanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.If explanted; give date: n/a (not applicable).Lens was removed/replaced during the same procedure.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis intraocular lens (iol) back haptic kinked upon insertion.The lens had to be cut out and replaced with a backup lens during the same procedure.There is no additional surgical intervention performed and no injury reported.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation: yes section d9: return to manufacturer: 3/16/2021 section h3: device evaluated by manufacturer: yes device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during removal and replacement.Haptic stuck to the optic and haptic damage (loop material broke off but not detached from drill hole) were also observed, but can be attributed to handling and cannot be confirmed to be related to manufacturing.Based on the return condition of the lens, no further product evaluation can be performed.The complaint issue was confirmed, however, this can be attributed to handling and cannot be confirmed to be related to manufacturing and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|