SMITHS MEDICAL ASD, INC. CADD-LEGACY DUODOPA AMBULATORY INFUSION PUMP; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 1400
Device Problem Inaccurate Delivery (2339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device evaluation: one smiths medical cadd-legacy duodopa ambulatory infusion pump was returned for analysis in used condition.Visual inspection showed the entire device was worn; front housing was scratched, rear housing was broken, there was fluid ingression on downstream occlusion sensor and the lock was worn.Review of the event history log showed occurrences of "high pressure" alarm messages, but the reported problem was not found.The investigation found that the flow rate was normal while testing and the device passed functional testing.Based on the investigation, the complaint allegation was not confirmed.
 
Event Description
Information was received indicating that a smiths medical cadd-legacy duodopa ambulatory infusion pump may have been delivering inaccurately.It was reported that with the same dose the amount remaining in the cartridge was inconsistent; sometimes more and sometimes less.It was also reported that although the pump indicated "func" it did not seem to deliver, the patient's daughter reported not hearing the dose administer.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-LEGACY DUODOPA AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11416291
MDR Text Key234882437
Report Number3012307300-2021-01746
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 03/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1400
Device Catalogue Number21-1400-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2021
Was the Report Sent to FDA? No
Date Manufacturer Received02/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N