| Model Number |
1012272-20 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Angina (1710); Low Blood Pressure/ Hypotension (1914); Vascular Dissection (3160); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Visual and functional inspections were performed on the returned device.The reported balloon rupture was confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a heavily calcified lesion in the left anterior descending artery.The 2.5x20mm trek rx balloon dilatation catheter was prepared per the instructions for use without issues; however, the balloon ruptured during the fourth inflation at 6-8 atmospheres.The procedure was successfully completed with the deployment of a 2.5x28mm unspecified stent.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported patient effects of angina, hypotension and dissection are listed in the coronary dilatation catheters, trek rx and mini trek rx, instructions for use as known patient effects.The investigation determined the reported balloon rupture, additional treatment and hospitalization appear to be related to operational context.A conclusive cause for the reported patient effects of angina, hypotension, and dissection and the relationship to the device, if any, cannot be determined.Attachment: user medwatch report #5099477 b1, b2, h1: malfunction to serious injury.H6: health effect - impact code 2199 removed.Investigation conclusions code 4307 removed.
|
| |
|
Event or Problem Description
|
|
Additional information: user facility medwatch report received that states: a (b)(6) yo f undergoing pci to the proximal to mid diagonal branch complicated by balloon rutpure (abbott balloon trek 2.5x20mm ref number: 1012272-20 lot#: 00420g1 exp: 2023-03-31) during low-pressure angioplasty and subsequent dissection propagating proximally.This was successfully contained and treated with appointment of 2 overlapping xience sierra drug-eluting stents.Patient c/o chest pain and experienced hypotension, put in a couple stents to fix the problem.Patient was addmitted for overnight stay and discharged on (b)(6) 2021 at her baseline.Fda safety report id #(b)(4).
|
| |
|