• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems Battery Problem (2885); Therapeutic or Diagnostic Output Failure (3023); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Event Description
Information was received via customer feedback stating the v60 device unexpectedly shut down, resulting in premature cessation of therapy delivery.The service engineer (se) inspected the device.The se found upon inspection that the unit powers on and cycles, but gives a ¿check vent battery failed¿ alarm message at start up.The se found the battery was heavily discharged and would not recharge.The se reviewed the ventilator diagnostic logs and noted that a significant timeframe from 1/28/2021 11:43 pm to 1/29/2021 7:18 am the unit ran on battery power the whole time until it finally shut off and power was restored again on 1/292021 7:41 am.It was also noted that the alarm volume was set at 2 during this run time.The customer biomedical engineer stated to believe this unit was plugged into a gfci (ground fault circuit interrupt) outlet that was tripped prior to use causing the unit to operate on battery only.The device was in clinical and therapeutic use at the time of the event.No report of harm or injury has been noted, nor provision of medical intervention to prevent permanent or life-threatening injury.The customer biomedical engineer replaced the battery and the unit was checked overall, run in tested, cleaned, functionally tested and no abnormality was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key11418514
MDR Text Key244197376
Report Number2031642-2021-00843
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public(01)00884838020054
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/04/2021
Initial Date FDA Received03/04/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-