Model Number TMT4B11
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
Date of Event 08/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Patient weight is not provided / unknown.Additional 510(k) numbers are k113753 and k112160.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
Doctor reports that the tmt4b11 implant was removed due to peri-implantitis.Bone loss, abscess, inflammation and apical perforation of the buccal cortical bone is also reported.Tooth site # 12.The osteotomy and defect were thoroughly debrided, a second fixture was paced along with allograft bone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP TM 4.1MM MTX FULL, 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key11423858
Report Number0002023141-2021-00549
Device Sequence Number19577976
Product Code DZE
UDI-Device Identifier00889024018945
UDI-Public(01)00889024018945(17)240824(10)1231594
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K132258
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/24/2024
Device Model NumberTMT4B11
Device Catalogue NumberTMT4B11
Device Lot Number1231594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2021
Initial Report FDA Received Date03/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-