| Model Number |
6379 |
| Medical Device Problem Codes |
No Display/Image (1183); Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported complaint was confirmed.The srt replaced the flow board.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
The service repair technician (srt) reported that during testing of the device at the service center, the unit did not display the flow board software version, only dashes.There was no patient involvement.
|
| |
|