• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRAXX NEPHROSTOMY BALLOON; LJE CATHETER, NEPHROSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRAXX NEPHROSTOMY BALLOON; LJE CATHETER, NEPHROSTOMY Back to Search Results
Model Number G46033
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Occupation: other non-healthcare professional: purchasing.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, during a percutaneous nephrolithotomy (pcnl) and using a ultraxx nephrostomy balloon, leakage from the balloon body occurred.The physician was inflating the device with water, and around 15atm the device began to leak.This occurred prior to use on the patient.Another device of the same type was used to complete the procedure.No section of the device remained inside the patient's body.The patient did not require any additional procedures due to this occurrence.The complainant has not reported any adverse effects on the patient due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.Correction- b1, h1: on return of device (b)(6) 2021, the balloon was inflated to 20 atm and was able to hold pressure at 20 atm.The bonds were strong and intact.A visual exam noted no leaks in the balloon or catheter.There is no evidence to suggest that the observed device condition would be likely to cause or contribute to a death or a serious injury if the failure were to recur.Furthermore, there is no evidence that the device caused or contributed to a serious injury, as no life-threatening or permanently impairing injury took place, nor was intervention taken as a result of the device failure which would be required to prevent permanent impairment or damage to the patient.As such, the event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury, or reportable product malfunction.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAXX NEPHROSTOMY BALLOON
Common Device Name
LJE CATHETER, NEPHROSTOMY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key11430118
Report Number1820334-2021-00901
Device Sequence Number17088049
Product Code LJE
UDI-Device Identifier00827002460331
UDI-Public(01)00827002460331(17)230727(10)13329087
Combination Product (Y/N)N
PMA/510(K) Number
K024050
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/27/2023
Device Model NumberG46033
Device Catalogue NumberUNB-6-15
Device Lot Number13329087
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2021
Initial Date Received by Manufacturer 02/26/2021
Supplement Date Received by Manufacturer03/25/2021
Initial Report FDA Received Date03/07/2021
Supplement Report FDA Received Date04/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-