MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. DORADO PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number DR135104
Device Problem Deflation Problem (1149)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 02/11/2021
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 05/2023).
 
Event Description
It was reported that during an angioplasty procedure, the pta balloon allegedly could not be fully deflated.It was further reported that a percutaneous needle stick was used to deflate the balloon.The current patient status is unknown.
 
Event Description
It was reported that during an angioplasty procedure, the pta balloon allegedly could not be fully deflated.It was further reported that a percutaneous needle stick was used to deflate the balloon.The current patient status is unknown.
 
Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the deflation problem could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.Expiry date: 05/2023 section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA DILATATION CATHETER
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11430963
MDR Text Key238356268
Report Number2020394-2021-00456
Device Sequence Number1
Product Code LIT
UDI-Device Identifier00801741059568
UDI-Public(01)00801741059568
Combination Product (y/n)N
PMA/PMN Number
K072283
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDR135104
Device Catalogue NumberDR135104
Device Lot Number93SE0021
Was Device Available for Evaluation? No
Date Manufacturer Received04/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-