• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE RX PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE RX PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U41502H10RX
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date: 06/2023.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a target lesion below the knee, the pta balloon allegedly ruptured at nominal pressure.It was further reported that another device was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE RX PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11431402
Report Number2020394-2021-00458
Device Sequence Number16954864
Product Code LIT
UDI-Device Identifier00801741056567
UDI-Public(01)00801741056567
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU41502H10RX
Device Catalogue NumberU41502H10RX
Device Lot NumberCMEU0174
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/12/2021
Initial Report FDA Received Date03/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-