MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 035 PTA CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U357574
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation; however, photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 08/2023).
 
Event or Problem Description
It was reported that during preparation of an angioplasty procedure, the pta balloon allegedly had foreign material inside the package.There was no patient contact.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: a sample was retuned for evaluation and no anomalies were noted to the device.The returned device was removed from the package but no signs of the foreign material was noted during the evaluation.No further functionality testing was conducted due to the nature of the complaint.One photo was reviewed by quality engineer.The photo shows a hair like foreign material found over the outer surface of the inflation luer, no other anomalies noted.Therefore, based on the photo review, the reported failure foreign material can be confirmed.However, the investigation will be inconclusive for the reported device contamination as there was no signs of the foreign material noted during evaluation although the photo attached was noted for the foreign material.A definitive root cause for the alleged foreign material could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4(expiry date: 08/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during preparation of an angioplasty procedure, the pta balloon allegedly had foreign material inside the package.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 035 PTA CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11432044
Report Number2020394-2021-00469
Device Sequence Number3287407
Product Code LIT
UDI-Device Identifier00801741093456
UDI-Public(01)00801741093456
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K142261
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU357574
Device Catalogue NumberU357574
Device Lot NumberCMEW0311
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2021
Initial Date Received by Manufacturer 02/11/2021
Supplement Date Received by Manufacturer06/11/2021
Initial Report FDA Received Date03/08/2021
Supplement Report FDA Received Date06/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-