| Model Number |
24715 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first date of the month of the aware date as no event date was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 5.0x150x150-hybrid sterling balloon catheter was advanced for dilation; however, during inflation at 12 atmospheres, the balloon ruptured.No known patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event updated.B5 describe event or problem updated.E1 initial reporter title updated.E1 initial reporter last name updated.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 5.0x150x150-hybrid sterling balloon catheter was advanced for dilation; however, during inflation at 12 atmospheres, the balloon ruptured.No known patient complciations were reported.It was further reported that the 60% stenosed target lesion was located in the moderately calcified and non-tortuous popliteal artery.Previously, it was reported that the balloon ruptured at 12 atmospheres.Now, it is reported that it ruptured at 13 atmospheres.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: the device was returned for analysis.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed a longitudinal tear in the balloon 38mm from the tip and is approximately 48mm long.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.A 5.0x150x150-hybrid sterling balloon catheter was advanced for dilation; however, during inflation at 12 atmospheres, the balloon ruptured.No known patient complciations were reported.It was further reported that the 60% stenosed target lesion was located in the moderately calcified and non-tortuous popliteal artery.Previously, it was reported that the balloon ruptured at 12 atmospheres.Now, it is reported that it ruptured at 13 atmospheres.
|
| |
|