ETHICON ENDO-SURGERY, LLC. ENSEAL G2 CURVED JAW 25CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLG2C25
Medical Device Problem Codes Difficult to Open or Close (2921); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: u93g1u.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the nslg2c25 device was received with the top of the i-blade fractured and slightly lifted and the upper jaw detached returned.The device was connected to the generator and it was recognized.Because the i blade was damaged not all functional testing could be performed with the generator.The condition of the blade prevented the functionality of the jaw.The jaw was unable to cycle open and close.It should be noted that product failure is multifactorial.A probable cause of the damage to the i blade could be not allowing thick and fibrous tissues to denature prior to i-blade advancement.As described in the ifu: do not use excessive force on the closing handle to close the jaws; grasp only as much tissue as will fit between the jaws where the current will pass.Greater amounts of tissue require more closing handle force.Excessive force could damage the device.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported that, during a laparoscopic partial hepatectomy, the handle was stiff and became not to close the jaws.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 CURVED JAW 25CM
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
MDR Report Key11435020
Report Number3005075853-2021-01269
Device Sequence Number3539463
Product Code GEI
UDI-Device Identifier10705036014133
UDI-Public10705036014133
Combination Product (Y/N)N
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberNSLG2C25
Device Catalogue NumberNSLG2C25
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/11/2021
Initial Report FDA Received Date03/08/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
-
-