• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA 100; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA 100; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARDLCA219001A
Device Problems Crack (1135); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.Device not returned to manufacturer.
 
Event Description
On 26th february, 2021 getinge became aware of an issue with lucea 100 surgical light.As it was stated, the covers were cracked and particles were missing.There was no injury reported, however we decided to report the issue in abundance of caution as any particles falling off into steriel field or during procedure may cause contamination.
 
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.
 
Event Description
Manufacturer reference number (b)(4).
 
Manufacturer Narrative
Getinge became aware of an issue with lucea 100 surgical light.As it was stated, the covers were cracked and particles were missing.There was no injury reported, however we decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.It was established that when the event occurred, the surgical light did not meet its specification as particles cover were missing, and it contributed to event.None of the provided information indicate that upon the event occurrence the device was being used for patient treatment.When reviewing similar reportable events for the same device type, we have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.If described failure occurs, the user can visually detect it during the daily checks to be performed prior to each use, or during preventive maintenance.In this case, the user would contact a getinge representative to replace the defective cover of the affected device.For cleaning, the ifu warns the user to not use aggressive and abrasive products.During disinfection, it is prohibited to spray the disinfectant solution directly on the device and to use inappropriate disinfectants.We believe that overall the devices on the market are performing correctly.Given the circumstances we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 100
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key11435879
MDR Text Key239162079
Report Number9710055-2021-00086
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberARDLCA219001A
Device Catalogue NumberARDLCA219001A
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/20/2021
Patient Sequence Number1
-
-