• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG PLUG F/TIBIAL PLATEAU; KNEE ENDOPROSTHETICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG PLUG F/TIBIAL PLATEAU; KNEE ENDOPROSTHETICS Back to Search Results
Model Number NN260P
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Implant Pain (4561); Swelling/ Edema (4577)
Date of Event 07/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation results.As of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.The device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.Two further incidents have been filed with the same lot regarding implant loosening until today.A clear root cause conclusion cannot be drawn due to the non-availability of important information and the device itself.The conclusion from the root cause analysis is that a systematic device deviation is unlikely, since the received complaints are limited to certain health care institutions and their applicants.The participation of the applicants in this deviation scenario cannot be excluded by today.Based on the root cause analysis result no corrective actions have been carried out.As a preventive action we were in contact with the affected health care institutions and applicants in order to reduce the probability of recurrence.The complaints were submitted to aesculap ag as a summary report.In order to examine further actions, all cases from the summary report are processed as vigilance suspicion cases and are continuously monitored.
 
Event or Problem Description
It was reported that there was an issue with vega knee implant(s).It was reported on (b)(6) 2018.That as a result of having the product implanted, the patient has experienced patient experienced pain, swelling, difficulty walking, and loosening of the implant which resulted in a revision surgery.The primary surgery occurred on (b)(6) 2013, and the revision surgery occurred on (b)(6) 2015.All available information has been provided at this time; if additional information becomes available the complaint will be updated accordingly.The adverse event is filed under xc reference (b)(4).Involved components: nx053z (as vega ps tibial plateau ), nx031z (ps femur cemented f5n rt), nx121 (ps pe insert t2/t2+, 12mm), nn260p (peek plug f/ tibia), nx043 (universal patella p3), the cement used was unknown associated medwatches: 2916714-2020-00324, 2916714-2020-00325, 2916714-2020-00326, 2916714-2020-00328.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUG F/TIBIAL PLATEAU
Common Device Name
KNEE ENDOPROSTHETICS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key11436259
Report Number2916714-2020-00327
Device Sequence Number14472296
Product Code JWH
UDI-Device Identifier04046963711320
UDI-Public4046963711320
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/08/2021
Report Date (Section F) 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/01/2018
Device Model NumberNN260P
Device Catalogue NumberNN260P
Device Lot Number51927002
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/20/2018
Event Location Hospital
Date Report to Manufacturer07/20/2018
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA03/08/2021
-
-