• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BURR, STRAIGHT, 19.5MM DIAM. 2.0 MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BURR, STRAIGHT, 19.5MM DIAM. 2.0 MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number BURR, STRAIGHT, 19.5MM DIAM. 2.0 MM
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that the device broke during a hallus valgus foot arthroscopy surgery inside the patient.This break was sharp at the base of the cutter head.Further approaches had to be created on the patient to retrieve the head of the device.There was no harm or adverse event for patient, operator or third party reported.The surgery was finished successfully with a new device with the same part number.It was not necessary to do a second surgery.Update 26-feb-2021: the lot number 022206 was provided and was corrected in the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, STRAIGHT, 19.5MM DIAM. 2.0 MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key11436698
Report Number1220246-2021-02685
Device Sequence Number5170113
Product Code LXH
UDI-Device Identifier00888867028654
UDI-Public00888867028654
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBURR, STRAIGHT, 19.5MM DIAM. 2.0 MM
Device Catalogue NumberAR-300-B201
Device Lot Number22206
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2021
Initial Report FDA Received Date03/08/2021
Patient Sequence Number1
-
-