• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS11.5-65B250
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Osteolysis (2377)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not been returned for evaluation.Root cause is unable to be determined at this time.Device labeling: metallic implants can loosen, fracture, corrode, migrate or cause pain.
 
Event or Problem Description
Information was received that a revision procedure was performed on (b)(6) 2021.As per the reporter, the patient experienced a sudden increase in pain and bone abnormalities at the junction of the telescoping nail segments.Upon explant it was noticed that the nail was corroded.No patient injury was reported.
 
Event or Problem Description
Additional information was received via review of literature that the patient experienced swelling, osteolysis and periosteal reaction.
 
Additional Manufacturer Narrative
Literature citation: rolfing, jan duedal, et al (2021).Pain, osteolysis, and periosteal reaction are associated with the stryde limb lengthening nail: a nationwide cross-sectional study.Acta orthopaedica, volume 92.Web address to article: https://doi.Org/10.1080/17453674.2021.1903278.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key11439689
Report Number3006179046-2021-00167
Device Sequence Number17018449
Product Code HSB
UDI-Device Identifier00887517950826
UDI-Public887517950826
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K180503
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPS11.5-65B250
Device Lot Number9070217
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/08/2021
Supplement Date Received by Manufacturer03/24/2021
Initial Report FDA Received Date03/08/2021
Supplement Report FDA Received Date04/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age42 YR
Patient Weight60
-
-