• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC MULTIDEBRIDER HANDPIECE, ANGLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC MULTIDEBRIDER HANDPIECE, ANGLED Back to Search Results
Model Number MDHP100A
Medical Device Problem Codes No Device Output (1435); Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has not yet been received for evaluation.The cause of the issue cannot be determined at this time.If additional information becomes available this report will be supplemented accordingly.
 
Event or Problem Description
It was reported that the handpiece is not functioning as intended.The device showed error message ¿e12 error¿.The issue occurred during an unspecified procedure.There was no patient harm or injury reported.No user injury reported.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The original equipment manufacturer (oem) performed a device history record review and no abnormalities were noted.The device was not returned for physical evaluation, however an investigation was completed by the oem and determine that the root cause of the failure reported can be attributed to a short circuit in the handpiece cable.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIDEBRIDER HANDPIECE, ANGLED
Common Device Name
MULTIDEBRIDER HANDPIECE, ANGLED
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key11440219
Report Number1037007-2021-00024
Device Sequence Number19713760
Product Code GEI
UDI-Device Identifier00821925027749
UDI-Public00821925027749
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K123429
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMDHP100A
Device Catalogue NumberMDHP100A
Device Lot NumberAV864322
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/11/2021
Supplement Date Received by Manufacturer04/06/2021
Initial Report FDA Received Date03/08/2021
Supplement Report FDA Received Date04/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-