Model Number 27093 |
Device Problem
Complete Loss of Power (4015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Resmed has requested for the device to be returned so that an engineering investigation could be performed.The device has not been returned, therefore resmed is unable to confirm the alleged malfunction at this time.(b)(4).
|
|
Event Description
|
It was reported to resmed that an astral device displayed a total power failure alarm and an error message related to a battery charger fault.There was no patient harm or serious injury reported as a result of this incident.
|
|
Manufacturer Narrative
|
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported total power failure alarm and revealed an error message (sf180) related to a battery charger fault.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the sf180 was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
|
|
Event Description
|
It was reported to resmed that an astral device displayed a total power failure alarm and an error message related to a battery charger fault.There was no patient harm or serious injury reported as a result of this incident.
|
|