• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - FUKUDA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - FUKUDA Back to Search Results
Model Number 27093
Device Problem Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2021
Event Type  malfunction  
Manufacturer Narrative
Resmed has requested for the device to be returned so that an engineering investigation could be performed.The device has not been returned, therefore resmed is unable to confirm the alleged malfunction at this time.(b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a total power failure alarm and an error message related to a battery charger fault.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported total power failure alarm and revealed an error message (sf180) related to a battery charger fault.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the sf180 was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference #: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a total power failure alarm and an error message related to a battery charger fault.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - FUKUDA
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key11440961
MDR Text Key238607021
Report Number3007573469-2021-00369
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Type of Report Initial,Followup
Report Date 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27093
Device Catalogue Number27093
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/21/2021
Distributor Facility Aware Date05/27/2021
Device Age39 MO
Event Location Hospital
Date Report to Manufacturer06/21/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/08/2021
Supplement Dates Manufacturer Received05/27/2021
Supplement Dates FDA Received06/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-