| Model Number |
ETUF3214C102E |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Obstruction/Occlusion (2422)
|
| Date of Event |
02/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: other relevant device(s) are: product id: e tlw1610c124e, serial/lot #: (b)(4), ubd: 15-mar-2019, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii aui stent graft system was implanted in the patient for the endovascular treatment of a 55 mm abdominal aortic aneurysm.It was reported that during a ct approximately 45 months post index procedure, it was noted that there was no blood flow through the aui and the limb.A thrombosed stent graft was noticed.A stent graft was implanted on the next day.The event is resolved and the blood flow was restored.As per the physician the cause of the event can not be determined.No additional clinical sequalae were provided and the patient will be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis the reported stent graft occlusion was confirmed on the films provided; however, the cause of the event could not be determined.Lack of pre-implant ct¿s did not allow for a thorough assessment of the pre-implant anatomy.Post-implant ct¿s were not available for review and the stent graft in-vivo configuration could not be evaluated.Additional contrast injection with the pig tail catheter positioned further distally in the stent graft was not performed for further analysis of the event.It is unknown if anatomical considerations that led to the implantation of an aui instead of a bifurcate stent graft may have contributed to the observed occlusive event.Other possible contributors to vessel occlusion include an unknown coagulopathy that is now presenting or a previous history of pad leading to poor distal run-off.Analysis of the returned films did not reveal any obvious out of specification stent graft integrity issues.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
B5.Additional information received; it was reported the physician was surprised by the occlusion, as there was nothing indicating it would occur and did not know the cause.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|