Model Number 68-750
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: ¿deflation¿.
 
Event or Problem Description
Healthcare professional reported left side deflation.Device remains implanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: one curved opening.A microscopic analysis and leak test were performed which identified one sharp opening.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: one sharp opening in the side valve bond edge assessed as unidentified (tear) opening.
 
Event or Problem Description
Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
MDR Report Key11445855
Report Number9617229-2021-02824
Device Sequence Number3523778
Product Code FWM
UDI-Device Identifier10888628000339
UDI-Public10888628000339
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/02/2019
Device Model Number68-750
Device Catalogue Number68-750
Device Lot Number2710491
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2021
Initial Date Received by Manufacturer 02/09/2021
Supplement Date Received by Manufacturer03/16/2021
Initial Report FDA Received Date03/09/2021
Supplement Report FDA Received Date04/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
-
-