| Model Number |
68-750 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/08/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: ¿deflation¿.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: one curved opening.A microscopic analysis and leak test were performed which identified one sharp opening.Fill inspection was performed and identified no blockage in the valve.Based on the device analysis the final assessment is: one sharp opening in the side valve bond edge assessed as unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|