• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. ALL POLY PATELLA CEMENTED 32 MM DIAMETER; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. ALL POLY PATELLA CEMENTED 32 MM DIAMETER; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 02/22/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2021-00050, 3007963827-2021-00051, 3007963827-2021-00052.Concomitant medical devices: femur cemented cruciate retaining (cr) narrow left size 9 catalog#: 42502006601 lot#: 64468334.Articular surface medial congruent (mc) left 12 mm height catalog#: 42512100812 lot#: 64421121.Natural tibia cemented 5 degree stemmed left size e catalog#:42532007101 lot#: 64402683.Report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that patient underwent a left knee revision approximately one year post implantation due to pain, swelling, ligament laxity, and swelling.During the revision, the tibial insert was found to be overly worn for the timeframe it was implanted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: immediate post-op images demonstrate anatomic alignment of the arthroplasty implants without abnormality.Pre-revision images demonstrate no interval change in implant position, implant loosening, or malalignment.Overall fit and alignment are normal and maintained on both image sets.Bone quality is osteopenic.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALL POLY PATELLA CEMENTED 32 MM DIAMETER
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
MDR Report Key11445897
Report Number0002648920-2021-00050
Device Sequence Number14292634
Product Code JWH
UDI-Device Identifier00889024247635
UDI-Public(01)00889024247635(17)270831(10)64509513
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42540000032
Device Lot Number64509513
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/22/2021
Supplement Date Received by Manufacturer05/07/2021
Initial Report FDA Received Date03/09/2021
Supplement Report FDA Received Date05/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age52 YR
Patient Weight123
-
-