• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER K2 EDTA (K2E) 10.8MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER K2 EDTA (K2E) 10.8MG BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367899
Medical Device Problem Codes Device Markings/Labelling Problem (2911); Missing Information (4053)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported the bd vacutainer® k2 edta (k2e) 10.8mg blood collection tubes experienced label lift off/partial peel off.The following information was provided by the initial reporter.The customer stated: "as i stated on the phone, it may be a label issue and not a tube issue.We were trying to determine the root cause and wondering if there was any new type of coating on the bd tubes.So far the new label we used today seems promising, with only one label lifting off of 52 tubes.".
 
Additional Manufacturer Narrative
H6: investigation summary: bd had received 1 samples and 1 photos for investigation.The photos were reviewed and the indicated failure mode for label lift with the incident lot was not observed.The samples were reviewed and the indicated failure mode for label lift with the incident lot was not observed.Additionally, 100 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to label lift as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.Bd was not able to identifiy a root cause for the indicated failure mode.
 
Event or Problem Description
It was reported the bd vacutainer® k2 edta (k2e) 10.8mg blood collection tubes experienced label lift off/partial peel off.The following information was provided by the initial reporter.The customer stated: "as i stated on the phone, it may be a label issue and not a tube issue.We were trying to determine the root cause and wondering if there was any new type of coating on the bd tubes.So far the new label we used today seems promising, with only one label lifting off of 52 tubes.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER K2 EDTA (K2E) 10.8MG BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
MDR Report Key11446208
Report Number1917413-2021-00167
Device Sequence Number4048563
Product Code JKA
UDI-Device Identifier50382903678995
UDI-Public50382903678995
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
BK980011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/31/2021
Device Model Number367899
Device Catalogue Number367899
Device Lot Number0133382
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2021
Initial Date Received by Manufacturer 02/17/2021
Supplement Date Received by Manufacturer03/14/2021
Initial Report FDA Received Date03/09/2021
Supplement Report FDA Received Date03/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-