WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA HELICAL BLADE 105MM STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
|
Back to Search Results |
|
| Model Number |
04.038.305S |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2021 follow up x-rays showed that a trochanteric fixation nail advanced (tfna) helical blade had cut through the acetabulum and entered the pelvis.The blade appeared free in rotation on the post op x-rays; an angio ct was preformed and no vessel damage was noted.The tfna was removed and the helical blade was extracted on (b)(6) 2021.The locking mechanism was not engaged and during the removal of the nail, the helical blade was free in rotation.The revision operation was completed successfully.Concomitant device: unk - screw: (part# unknown; lot# unknown; quantity: unknown).This report is for one (1) tfna helical blade 105mm sterile.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4) depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: investigation summary customer quality investigation: the complaint device was not received for investigation.The following investigation is based on the images provided in the eight attachments uploaded by feb 15, 2021.The images were reviewed, and the complaint condition was confirmed as the helical blade can be seen to have migrated further into the patient.Since the device was not returned, a dimensional inspection and a functional test were not able to be performed.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device history lot manufacturing location: elmira / packaged, sterilized and released by: monument manufacturing date: sep 10, 2018, expiration date: jul 31, 2028, part number: 04.038.300s, tfna fenestrated helical blade 100mm -sterile.Lot number: h719619 (sterile), lot quantity: 48.Work order traveler met all inspection acceptance criteria.Inspection sheet, ns065693 met all inspection acceptance criteria.Packaging label log (pll) lppf, lmd was reviewed and determined to be conforming.Packaging bom was reviewed and determined to be conforming with no deviations to normal packaging identified.Scn 15459 supplied by sterigenics was reviewed and determined to be conforming.This lot met all visual, sterility and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device history review feb 15, 2021: dhr reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: visual inspection: the tfna helical blade 105mm sterile (p/n: 04.038.305s, lot number: h719619) was received at us cq.Visual inspection of the complaint device showed no external damage.Review of the attached x-rays showed that the complaint device had migrated from its initial position.The returned mating device had damage to its locking prong subcomponent that resulted in an inability to hold the complaint device in place.Functional test: a functional assessment was performed on the complaint device.The complaint device could assemble with the returned mating device but would not hold its position.The complaint was able to be replicated.Document/specification review: no design issues or discrepancies were identified.Complaint confirmed? yes.Investigation conclusion: this complaint is confirmed as the x-rays showed the complaint device having migrated.However, this is likely due to damage on the mating device and not due to the complaint device.No definitive root cause could be determined based on the provided information.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|