• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SERV REPL ASCOPE A/C HD 30 DEG 4MM; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SERV REPL ASCOPE A/C HD 30 DEG 4MM; ARTHROSCOPE Back to Search Results
Model Number 72202087S
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the scope presented an unknown malfunction.No case reported; therefore, there was no patient involvement.Results of investigation have concluded that this unit had a distal lens is cracked which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3,h6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection found a deep dented needle, distal tip and fiber damage, and the distal lens is cracked.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat event.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERV REPL ASCOPE A/C HD 30 DEG 4MM
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key11447330
Report Number3003604053-2021-00071
Device Sequence Number19777264
Product Code HRX
UDI-Device Identifier03596010644534
UDI-Public03596010644534
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K962075
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202087S
Device Catalogue Number72202087S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Initial Date Received by Manufacturer 02/18/2021
Supplement Date Received by Manufacturer04/15/2021
Initial Report FDA Received Date03/09/2021
Supplement Report FDA Received Date04/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-