U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number ZCU300
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight, ethnicity: unknown, not provided.(b)(4).All pertinent information available to (b)(6) has been submitted.
 
Event or Problem Description
It was reported that a tecnis toric ii intraocular lens (iol) had mechanical issues due to incorrect diopter used.The lens was later explanted and replaced with a model zcu375, 21.0 diopter iol.There was no additional surgical intervention required and no injury was reported.Reportedly, the patient is doing good post-exchange.No further information provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Section d9: return to manufacturer: 3/16/2021.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during explant.A detached haptic was observed, which can be attributed to handling and cannot be confirmed to be related to manufacturing.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation, there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
MDR Report Key11448345
Report Number9614546-2021-07093
Device Sequence Number3618701
Product Code HQL
UDI-Device Identifier05050474660168
UDI-Public(01)05050474660168(17)240920
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/20/2024
Device Model NumberZCU300
Device Catalogue NumberZCU300U205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/10/2021
Supplement Date Received by Manufacturer04/04/2021
Initial Report FDA Received Date03/09/2021
Supplement Report FDA Received Date04/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
-
-