• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CONTROL UNIT DYONICS 25; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CONTROL UNIT DYONICS 25; ARTHROSCOPE Back to Search Results
Model Number 7211010
Device Problems Increase in Pressure (1491); Free or Unrestricted Flow (2945)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event Description
It was reported that during a knee arthroscopy while using the control unit dyonics 25 the pump would shoot up in pressure and not hold at the designated setting.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A complaint history review concluded this was a repeat issue.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTROL UNIT DYONICS 25
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11448637
MDR Text Key239439573
Report Number1643264-2021-00934
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010543554
UDI-Public03596010543554
Combination Product (y/n)N
PMA/PMN Number
K171794
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 04/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7211010
Device Catalogue Number7211010
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/16/2021
Initial Date FDA Received03/09/2021
Supplement Dates Manufacturer Received03/30/2021
Supplement Dates FDA Received04/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-