• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR RX HERCULINK ELITE STENT SYSTEM; STENT, RENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR RX HERCULINK ELITE STENT SYSTEM; STENT, RENAL Back to Search Results
Model Number 1011505-15
Medical Device Problem Codes Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/27/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up will be submitted with all relevant information.Sus event #mw5098900.
 
Event or Problem Description
User facility medwatch report received that states: the patient had left subclavian stenosis involving the ostium of the left vertebral artery.To induce retrograde vertebral flow throughout the case, the operating physician opted to place a moma device within the proximal subclavian artery.The distal balloon was removed and only the proximal occlusion was performed.After stabilizing the moma device, the proximal balloon was inflated in the very proximal left subclavian ensuring retrograde flow to the left vertebral.The left subclavian artery was successfully stented with a rx herculink elite 7 x 15 stent.Upon removal of the initial herculink stent balloon the balloon ruptured and a very small portion of the distal balloon may have been sheared off by moma device.However, no evidence of embolization was noted.The patient tolerated the procedure well without complications and was notified by the physician.Fda safety report id # (b)(4).
 
Additional Manufacturer Narrative
Product performance engineering reviewed the incident information; however, the product was not returned to abbott vascular for analysis.Additional information and/or return of the device may have further aided the analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.Balloon material ruptures can be affected by numerous factors such as balloon damage during processing of the balloon material, materials, inflation technique, inflation over rated burst pressure, interactions with other devices, anatomical conditions, a previously implanted stent or insufficient preparation prior to use.The device was prepped prior to use without any leak or ruptures noted, which would suggest that the device was not damaged prior to use.Based on the limited information provided, the investigation was unable to determine a conclusive cause for the reported balloon rupture.In this case, it is possible that the balloon became damaged against the stent implant resulting in the reported balloon rupture; however, this could not be confirmed.Manipulation of the device during retraction likely caused the balloon separation in the anatomy.There is no indication of a product quality issue with respect to manufacture, design or labeling.The viatrac 14 balloon referenced in b5 is filed under a separate medwatch report number.D9, h3 - the device was initially reported as returning for analysis, but the device was not returned.
 
Event or Problem Description
Subsequent to the initially filed report, the device returned to abbott with the user facility medwatch report was a 7x20 mm viatrac 14 balloon dilatation catheter.Analysis of the device observed a balloon rupture.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX HERCULINK ELITE STENT SYSTEM
Common Device Name
STENT, RENAL
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11448806
Report Number2024168-2021-01883
Device Sequence Number3497161
Product Code NIN
UDI-Device Identifier08717648078378
UDI-Public08717648078378
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P110001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model Number1011505-15
Device Catalogue Number1011505-15
Device Lot Number0051461
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/15/2021
Supplement Date Received by Manufacturer04/29/2021
Initial Report FDA Received Date03/09/2021
Supplement Report FDA Received Date05/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MOMA EMBOLIC PROTECTION DEVICE
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
-
-